REOTEMP Adjustable Angle Digital Thermometer/Transmitter
|
|
Choose Your Model
| 4-20 mA OUTPUT LOCATION: |
|
| Process Connection: |
|
| Temperature Range: |
|
Options
* denotes required field
Stem Diameter*
Stem Length*
|
REOTEMP Adjustable Angle Digital Thermometer/Transmitter
The Reotemp Adjustable Angle Digital Thermometer/Transmitter is a durable high accuracy thermometer with digital display. It features an IP67/NEMA 4X enclosure, 5 year battery life and optional transmitter output (Note: 4-20mA models are loop powered 9-36VDC).
- High Accuracy 1,000Ω RTD (Class A)
- Water Resistant IP67/NEMA 4X
- 5 Year Battery Life (4-20mA Models are Loop Powered)
- 4-20 mA Output Available (Loop Powered 9-36VDC)
- Low or High Temperature Range Models

Specifications
| Accuracy
|
± ( 0.72 + 0.002 x | t-32 | ) °F,± ( 0.40 + 0.002 x | t | ) °C, where t = temperature. Example: At 32°F accuracy is ± 0.72 °F.
|
| Sensing Element
|
RTD, Type Pt1000Ω, Class A
|
| Temperature Ranges
|
-58°F to 392°F (-50°C to 200°C) or -328°F to 1112°F (-200°C to 600°C)
|
| Output (Optional)
|
4-20 mA, M12 Connector, Loop Powered 9-35 VDC
|
| Refresh Rate
|
3 Seconds
|
| Display
|
4-digit LCD, 1/2” Height
|
| Display Resolution
|
See Table 1
|
| RFI Effect
|
1% or Less Typical
|
| Temperature Effect
|
<0.01% FS/°C
|
| Ambient Temperature Range
|
32°F to 122°F (0°C to 50°C)
|
| Storage Temperature Range
|
-4°F to 158°F (-20°C to 70°C)
|
| Housing Material
|
Stainless Steel 316
|
| Lens
|
Plastic Polycarbonate (Standard) or Glass
|
| Probe Material
|
304 Stainless Steel
|
| Weight
|
12 oz., Varies by Configuration
|
| Environmental Protection
|
NEMA 4X/IP67
|
| Power
|
1 x 3.6V AA Battery (M12 is loop powered 9-36 VDC.)
|
| Battery Life
|
5 Years Minimum in Continuous Mode, (Note: 4-20mA Models are Loop Powered 9-36 VDC, No Battery Option)
|
| Electronic Display Temperature Limit
|
If the temperature of the electronics in the display housing exceeds 122°F (50°C), permanent damage to the display will occur. A remote mount is recommended for high or low temperature applications.
|
| Table 1: Display Resolution
|
| Model
|
Range
|
Decimal Point
|
Standard Temp,
Displays Decimal
|
-58°F to 392°F
(-50°C to 200°C )
|
Yes
|
Extended Temp,
No Decimal
|
-328°F to 1112°F
(-200°C to 600°C)
|
No
|