Additel AccuMac AM1760 and AM1762 Secondary SPRTs
|
Manufacturer: ADDITEL *Call for Pricing*
Choose Your Model
Model: |
|
Size: |
|
Connection: |
|
|
Additel AccuMac AM1760 and AM1762 Secondary SPRTs
- Reliable SPRT for Daily Use
- Rtpw Drift: <4mK after 100hrs at 661°C
- Short-term Stability : <0.002°C at 0.01°C
- Temperature Range: -200°C to 670°C
- Inconel™ Sheath to Withstand Harsh Environments
- Fully Meets the ITS-90 Criteria for Reference Thermometers
- ISO 17025 Accredited Calibration Included
The AM1760 series Secondary SPRTs provides our customers with reliable secondary standards that can be used daily in their labs. These SPRTs feature an accuracy of <0.006°C at 0.01°C, a short term stability of <0.002°C and a very low drift rate of less than 0.004°C after 100 hours at 661°C. Two different lengths of these SPRTs are available at either12 inches (305 mm) or 20 inches (508 mm).
The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving these SPRTs a high level of stability and repeatability in performance. A uniquely designed support structure and filling material provides excellent balance between the hysteresis effect, mechanical shock and thermal shock performance.
This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an ISO 17025 Accredited Calibration.
.
Optional Accessories:
9070- Smart connector for reference PRT for use with Additel temeprature products
9071- Connector adapter from Additel smart connector to 4-wire with gold plated spades
9072- Connector adapter from Additel smart connector to 4-wire with clamps
Note: Carrying Case Included (Excludes - Bend probes)
General Specifications |
|
|
AM1760 |
AM1762 |
AM1760-BEND |
AM1762-BEND |
Temperature Range
|
-200°C to 670°C
|
Nominal Resistance at 0°C
|
100 Ω
|
25 Ω
|
100 Ω
|
25 Ω
|
Temperature Coefficient
|
0.003925 Ω/Ω/°C
|
Calibrated Accuracy (k=2) [1][2]
|
±0.010°C at -196°C
|
±0.006°C at 0.01°C
|
±0.015°C at 420°C
|
±0.025°C at 661°C
|
Long Term Drift at 0.01°C [3]
|
<0.004°C at TPW after 100 hours at 661°C
|
Short Term Stability
|
<0.002°C
|
Thermal Shock
|
<0.002°C after 10 thermal cycles from minimum to maximum temperatures
|
Self-heating
|
0.0015°C at 0.5 mA current
|
0.0015°C at 1 mA
|
0.0015°C at 0.5 mA
|
0.0015°C at 1 mA
|
Response Time [4]
|
9 seconds
|
Measurement Current
|
0.5 mA
|
1 mA
|
0.5 mA
|
1 mA
|
Sensor Length
|
1.65 in (42 mm)
|
Sensor Location
|
0.2 in (5 mm) from tip
|
Insulation Resistance
|
>1000 MΩ at room temperature
|
Sheath Material
|
Inconel™
|
Sheath Dimensions
|
1760-20: 0.25 in (6.35 mm) (OD) x 20 in (508 mm) (L)
|
1762-20: 0.25 in (6.35 mm) (OD) x 20 in (508 mm) (L)
|
0.25 in (6.35 mm) (OD) x 9.75 in (247.65 mm) vertical x 4.5 in (114.3 mm) horizontal
|
1760-12: 0.25 in (6.35 mm) (OD) x 12 in (305 mm) (L)
|
1762-12: 0.25 in (6.35 mm) (OD) x 12 in (305 mm) (L)
|
External Leads
|
Teflon™ insulated copper wire, 4 leads, 6.5 feet (2 meters)
|
|
Handle Dimensions
|
0.59 in (15 mm) (OD) x 2.56 in (65 mm) (L)
|
Handle Temperature Range
|
-50°C to 180°C
|
|
[1] Includes calibration and 100 hour drift.
|
|
[2] Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges).
|
|
[3] Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
|
|
[4] For 63% response to step change in water moving at 1 meter per second.
|
|
|
Temperature Range: -200°C to 160°C
Accuracy: <0.05°C
Short-term Stability: <0.02°C
Transition Junction and Lead Wires Can Withstand the Full Range of the PRT's
Widely Used in Climate/Humidity Chambers and Freezers for Validation/Calibration
|
Temperature Range: -40°C to 160°C
Accuracy: <0.012°C at 0.01°C
Long Term Drift: <0.01°C
Short-term Stability: <0.007°C
Durable and Shock Resistant
Temperature Coefficient 0.003925ΩΩ/ Ω/ °C
Fully Meets the ITS-90 Criteria for Reference Thermometers
|
Affordable Reference Probe
Temperature Range: -200°C to 670°C
Accuracy: <0.012°C at 0.01°C
Short-term Stability: <0.007°C
Temperature Coefficient 0.003925 Ω/Ω/°C
Fully Meets the ITS-90 Criteria for Reference Thermometers
Inconel™ Sheath to Withstand
|
Temperature Range: -200°C to 670°C
Accuracy: <0.035°C
Long Term Drift: <0.03°C
Short-term Stability: <0.01°C
Durable and Shock Resistant
Temperature Coefficient 0.00385 Ω/ Ω/ °C
Inconel™ Sheath on AM1660/1640 Models
|
Affordable Reference Probe
Accuracy: <0.012°C at 0.01°C
Short-term Stability: <0.007°C
Temperature Range: -200°C to 420°C
Sheath Diameters Available in 2 Configurations: 1/4 inch and 3/16 inch
Inconel™ Sheath to Withstand Harsh Environments
Fully Me
|
Affordable Reference Standard
Type S
Short-term Stability: <0.2°C at 1084.62°C
Temperature Range: -196°C to 1300°C
|
|
|