Additel AccuMac AM1751 Secondary Reference PRT
|
Manufacturer: ADDITEL *Call for Pricing*
Choose Your Model
Size: |
|
Connection: |
|
|
Additel AccuMac AM1751 Secondary Reference PRT
- Affordable Reference Probe
- Temperature Range: -200°C to 670°C
- Accuracy: <0.012°C at 0.01°C
- Short-term Stability: <0.007°C
- Temperature Coefficient 0.003925 Ω/Ω/°C
- Fully Meets the ITS-90 Criteria for Reference Thermometers
- Inconel™ Sheath to Withstand Harsh Environments
- ISO 17025 Accredited Calibration Included
The AM1751 series of Secondary Reference PRTs offers affordable reference probes for precision temperature measurement and calibration both in the lab and in the field. These PRTs feature an accuracy of <0.012°C and a short term stability of <0.007°C.
To reach the best performance in stability and repeatability, the sensing element has been specially designed to protect the platinum sensing wire from contamination at high temperatures. A unique supporting structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard with an ISO 17025 Accredited Calibration Included.
Optional Accessories:
9070- Smart connector for reference PRT for use with Additel temeprature products
9071- Connector adapter from Additel smart connector to 4-wire with gold plated spades
9072- Connector adapter from Additel smart connector to 4-wire with clamps
Note: Carrying case included (Excludes Bend Probes)
General Specifications |
|
AM1751-12 |
AM1751-BEND |
Temperature Range
|
-200°C to 670°C
|
Nominal Resistance at 0.01°C
|
100 Ω
|
Temperature Coefficient
|
0.003925 Ω/Ω/°C
|
Calibrated Accuracy (k=2) [1][2]
|
±0.025°C at -40°C
|
±0.015°C at 0.01°C
|
±0.035°C at 420°C
|
±0.05°C at 661°C
|
Long Term Drift at 0.01°C [3]
|
<0.01°C at TPW after 100 hours at max temperature
|
Short Term Stability
|
<0.007°C
|
Thermal Shock
|
<0.005°C after 10 thermal cycles from minimum to maximum temperatures
|
Hysteresis
|
≤0.005°C
|
Self-heating
|
50 mW/°C
|
Response Time [4]
|
9 seconds
|
Measurement Current
|
0.5 mA or 1 mA
|
Sensor Length
|
1.26 in (32 mm)
|
Sensor Location
|
0.12 in (3 mm) from tip
|
Insulation Resistance
|
>1000 MΩ at room temperature
|
Sheath Material
|
Inconel™
|
Sheath Dimensions
|
0.25 in (6.35 mm) (OD)
|
0.25 in (6.35 mm) (OD)
|
X 12 in (305 mm) (L)
|
X 9.75 in (247.65 mm) vertical
|
|
X 4.5 in (114.3 mm) horizontal
|
External Leads
|
Teflon™ insulated copper wire, 4 leads, 6.5 feet (2 meters)
|
Handle Dimensions
|
0.59 in (15 mm) (OD)
|
X 2.56 in (65 mm) (L)
|
Handle Temperature Range
|
-50°C to 180°C
|
[1] Includes calibration and 100 hour drift.
|
[2] Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges).
|
[3] Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
|
[4] For 63% response to step change in water moving at 1 meter per second.
|
|
Temperature Range: -200°C to 160°C
Accuracy: <0.05°C
Short-term Stability: <0.02°C
Transition Junction and Lead Wires Can Withstand the Full Range of the PRT's
Widely Used in Climate/Humidity Chambers and Freezers for Validation/Calibration
|
Temperature Range: -40°C to 160°C
Accuracy: <0.012°C at 0.01°C
Long Term Drift: <0.01°C
Short-term Stability: <0.007°C
Durable and Shock Resistant
Temperature Coefficient 0.003925ΩΩ/ Ω/ °C
Fully Meets the ITS-90 Criteria for Reference Thermometers
|
Temperature Range: -200°C to 670°C
Accuracy: <0.035°C
Long Term Drift: <0.03°C
Short-term Stability: <0.01°C
Durable and Shock Resistant
Temperature Coefficient 0.00385 Ω/ Ω/ °C
Inconel™ Sheath on AM1660/1640 Models
|
Affordable Reference Probe
Accuracy: <0.012°C at 0.01°C
Short-term Stability: <0.007°C
Temperature Range: -200°C to 420°C
Sheath Diameters Available in 2 Configurations: 1/4 inch and 3/16 inch
Inconel™ Sheath to Withstand Harsh Environments
Fully Me
|
Reliable SPRT for Daily Use
Rtpw Drift: <4mK after 100hrs at 661°C
Short-term Stability : <0.002°C at 0.01°C
Temperature Range: -200°C to 670°C
Inconel™ Sheath to Withstand Harsh Environments
Fully Meets the ITS-90 Criteria for Reference Thermometers
|
Affordable Reference Standard
Type S
Short-term Stability: <0.2°C at 1084.62°C
Temperature Range: -196°C to 1300°C
|
|
|