Additel AccuMac AM1210 Type S Thermocouple
|
Manufacturer: ADDITEL *Call for Pricing*
Choose Your Model
Size: |
|
Junction: |
|
|
Additel AccuMac AM1210 Type S Thermocouple
- Affordable Reference Standard
- Type S
- Short-term Stability: <0.2°C at 1084.62°C
- Temperature Range: -196°C to 1300°C
The AM1210 Reference Standard Type S Thermocouple is made from reference grade platinum and platinum-rhodium alloy. It covers a temperature range from 0°C to 1300°C with a short term stability of <0.2°C all the way to the Freezing Point of Copper (1084.62°C). It is commonly used as a reference standard to calibrate industrial thermocouples.
All thermocouple wires and parts are specially cleaned and annealed before assembly. Every AM1210 thermocouple is fully annealed and tested again after assembly to meet the tolerance criteria as specified below:
E(tCu)=10.575 ±0.015
E(tAl)=5.860+0.37(E(tCu)-10.575) ±0.005
E(tZn)=3.447+0.18(E(tCu)-10.575) ±0.005
General Specifications |
|
M1210-20 |
AM1210-12 |
Temperature Range
|
0°C to 1300°C
|
Type
|
Type S: Platinum/10% Rhodium vs. Platinum
|
Long Term Drift*
|
<0.5°C at 1210°C after 1 year typical usage
|
Tolerance (mV)
|
E(tCu)=10.575 ±0.015
|
E(tAl)=5.860+0.37(E(tCu)-10.575) ±0.005
|
E(tZn)=3.447+0.18(E(tCu)-10.575) ±0.005
|
Short Term
|
<0.2°C at 1084.62°C
|
Diameter of Thermocouple Wire
|
0.02 in (0.5 mm)
|
Sheath Material
|
Alumina
|
|
Sheath Dimensions
|
0.236 in (6mm) (OD) x 19.68 in (500 mm) (L)
|
0.236 in (6mm) (OD) x 11.81 in (300 mm) (L)
|
Total TC Wire Length
|
47.25 in (1200 mm)
|
39.37 in (1000 mm)
|
Documentation**
|
Report of test with data
|
|
* Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
|
** For 17025 accredited calibration, please contact Additel.
|
Additel AM1210 Datasheet
|
Temperature Range: -200°C to 160°C
Accuracy: <0.05°C
Short-term Stability: <0.02°C
Transition Junction and Lead Wires Can Withstand the Full Range of the PRT's
Widely Used in Climate/Humidity Chambers and Freezers for Validation/Calibration
|
Temperature Range: -40°C to 160°C
Accuracy: <0.012°C at 0.01°C
Long Term Drift: <0.01°C
Short-term Stability: <0.007°C
Durable and Shock Resistant
Temperature Coefficient 0.003925ΩΩ/ Ω/ °C
Fully Meets the ITS-90 Criteria for Reference Thermometers
|
Affordable Reference Probe
Temperature Range: -200°C to 670°C
Accuracy: <0.012°C at 0.01°C
Short-term Stability: <0.007°C
Temperature Coefficient 0.003925 Ω/Ω/°C
Fully Meets the ITS-90 Criteria for Reference Thermometers
Inconel™ Sheath to Withstand
|
Temperature Range: -200°C to 670°C
Accuracy: <0.035°C
Long Term Drift: <0.03°C
Short-term Stability: <0.01°C
Durable and Shock Resistant
Temperature Coefficient 0.00385 Ω/ Ω/ °C
Inconel™ Sheath on AM1660/1640 Models
|
Affordable Reference Probe
Accuracy: <0.012°C at 0.01°C
Short-term Stability: <0.007°C
Temperature Range: -200°C to 420°C
Sheath Diameters Available in 2 Configurations: 1/4 inch and 3/16 inch
Inconel™ Sheath to Withstand Harsh Environments
Fully Me
|
Reliable SPRT for Daily Use
Rtpw Drift: <4mK after 100hrs at 661°C
Short-term Stability : <0.002°C at 0.01°C
Temperature Range: -200°C to 670°C
Inconel™ Sheath to Withstand Harsh Environments
Fully Meets the ITS-90 Criteria for Reference Thermometers
|
|
|