Additel ADT 919A High Pressure Test Pump
|
Choose Your Model
Pressure Port: |
|
Options
ADT 102 - Adapter and Fitting, male to various male and female connections (25 pc)
ADT 103 - Adapter and Fitting, male to various female hand-tight quick connection set (10 pc with carrying case)
Optional Pressure Manifold
|
Additel ADT 919A High Pressure Test Pump-14 to 2000 psi, pneumatic air pressure pump
OVERVIEW
The Additel ADT 919 High Pressure Test Pump is a hand operated pressure pump designed to generate pressure from -14psi (-0.95bar) to 2000psi (140bar). With a long lever, it just takes 30 seconds to reach 2000 psi (140 bar). A high-quality screw press is designed for fine pressure adjustment, with an adjustment resolution up to 10 Pa (0.1 mbar). A specially designed shut-off valve makes the pressure as stable as possible during calibration. A built-in gas-liquid isolator protects the pump from moisture and dirt to reduce the need for maintenance. The residual liquid in the pump introduced from devices under test will be pushed out and collected during pressure release. Two hand-tight connectors installed on the pump allow easy connecting and disconnecting to the test pump without the need for PTFE tape or wrenches. The ADT 919 is an ideal comparison test pump for pressure instruments calibration.
GENERAL FEATURES
• High Efficiency
Generate 2000 psi (140 bar) in 30 seconds
• High Resolution
10 Pa (0.1 mbar)
High-quality screw press for fine pressure adjustment
• Great Stability
A specially designed shut-off valve makes the pressure as stable as possible during calibration.
• Durable and Minimal Maintenance
Built-in gas-liquid isolator protects the pump from moisture and dirt
The residual liquid in the pump introduced from devices under test will be pushed out and collected during pressure release
• Hand-tight Quick Connectors
Allows easy connecting and disconnecting to the test pump without the need for PTFE tape or wrenches.
APPLICATIONS
The ADT 919 is designed to generate pressure from 95% vacuum to 2000 psi (140 bar) for the high pressure application requiring accurate pressure testing and calibration. It can be used in calibrating high pressure gauges, transmitters, or other pressure measuring instruments.
SPECIFICATIONS
• Media: Air.
• Generated Pressure Range:
14 psi (0.95 bar) vacuum to 2000 psi (140 bar) positive pressure
• Pressure Resolution
10 Pa (0.1 mbar)
• Material
Ram/adapters: SST
Body: SST/aluminum
Seals: Buna-N
• Connection
Hand-tight connectors for both test gauge and reference gauge.
Test Gauge Connection: 1/4" NPT female, 1/4" BSP female, M20X1.5 female
Reference Gauge Connection: 1/4" NPT female , 1/4" BSP female, M20X1.5 female
• Dimensions
Height: 7.00'' (178 mm);
Base: 21.26' (540 mm) x 10.63'' (270 mm)
• Weight: 16.4 lb (7.4 kg).
• Warranty: 1 year
• ORDERING INFORMATION:
• Accessories (included):
O-ring 20 pc