Fluke/Hart 5627A - Precision Industrial RTDs
|
Manufacturer: FLUKE/HART SCIENTIFIC
Mfr. Part Number: 5627A
Choose Your Model
Model: |
|
|
Fluke/Hart 5627A - Precision Industrial RTDs
Features
* Vibration and shock resistant
* 19 mm (3/4-inch) bend radius for increased durability
* NVLAP-accredited calibration included, lab code 200706-0
Specifications
Resistance
|
Nominal 100W
|
Temperature Coefficient
|
0.00385 W/W/ °C nominal
|
Temperature Range
|
–200 °C to 420 °C
(5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
|
Drift Rate (k=2)
|
± 0.04 °C at 0 °C after 100 hours at 420 °C
|
Sheath Material
|
316 Stainless Steel
|
Leads
|
Teflon™-insulated, nickel-plated stranded copper, 22 AWG
|
Termination
|
Specify. See Ordering Information.
|
Time Constant
|
Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
|
Bending Radius
|
Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Hart lab requires 20 cm [8 in] of unbent sheath to re-calibrate.)
|
Calibration
|
Includes manufacturer’s NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from –196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included.
|
Immersion
|
At least 100 mm (4 in) recommended
|
Calibrated Accuracy† (k=2)
|
± 0.026 °C at –196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C
|
Size
|
5627A-12: 305 mm x 6.35 mm (12 in x 1/4 in)
5627A-9: 229 mm x 4.7 mm (9 in x 3/16 in)
5627A-6: 152 mm x 4.7 mm (6 in x 3/16 in)
|
†Includes calibration uncertainty and 100 hr drift.
|
When buying a PRT, performance isn’t the only criterion you need to look at. The real issues are price-to-accuracy and price-to-durability ratios.
5627A probes have a temperature range up to 420 °C and an accuracy as good as ± 0.025 °C. They come in three different lengths. (Both six- and nine-inch models cover –200 °C to 300 °C.) Each instrument is shipped with its ITS-90 coefficients and a calibration table in 1 °C increments.
One of the best features of this sensor is that it conforms to the standard 385 curve, letting you use your DIN/IEC RTD meters fully. Why use a probe that’s less accurate than your meter?
The 5627A is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. (Bend, if any, should be specified at time of order.)
Six- and nine-inch 5627As are calibrated at –196 °C, –38 °C, 0 °C, 200 °C, and 300 °C. For 12-inch versions the point at 300 °C is replaced by a calibration point at 420 °C.
Each probe is individually calibrated and includes a NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.
This probe is an excellent value. It has the price-to-accuracy and price-to-durability ratios you should demand in every PRT you buy!
Ordering Information
5627A-6-X Secondary PRT, 152 mm x 4.7 mm (6 x 3/16 in), –200 °C to 300 °C
5627A-9-X Secondary PRT, 229 mm x 4.7 mm (9 x 3/16 in), –200 °C to 300 °C
5627A-12-X Secondary PRT, 305 mm x 6.35 mm (12 x 1/4 in), –200 °C to 420 °C
|
Case, Probe Carrying. Protect your PRT's! Keep them in a padded case between uses.
| | |
|
|