Kanomax Airflow Transducer Model 6332
|
|
Item Is In Stock
|
|
|
|
|

Product Line Quality Control PCB Temperature Monitoring
Quality Control monitoring and controlling Combined with thermo graphic system,
air velocity and distribution during drying/ measure air velocity across the board
shaping process. to evaluate the cooling efficiency.
(Photo provided by NEC San-ei Instruments, Ltd.)

Air Velocity Monitor for Local Residential Air Ventilation Monitoring
Exhaust Ventilation and FFU Ensure ventilation system efficiency by
(Fan Filter Unit) actively monitoring system balance
Prevent exposure of hazardous and controlling airflow.
materials to workers and prevent
leakage by thorough monitoring
and control of chamber airflow.
|
Features
- 10 interchangeable probe options
- Probes quickly and easily attach/detach
- Digital Display Option (model 6332D)
- Selectable Output (Output can be set to either voltage (0 to 5V) or current (4 to 20mA))
|
Main Unit Specifications
|
Product Name
|
Airflow Transducer (with/without Display)
|
|
Model
|
6332 / 6332D
|
|
Measuring Object
|
Clean air and compatible, non-combustible gases
|
|
Measuring Range
|
Maximum range 0.1 ~ 50 m/s (Range dependant on probe type*)
|
|
Measuring Accuracy
|
± (3% of reading + 0.1) m/s
|
|
Temperature Compensation Range
|
|
0.1-4.99m/s
|
5.00-9.99m/s
|
10.0-24.9m/s
|
25.0-50.0m/s
|
|
5~40°C
|
±0.25m/s
|
±0.50m/s
|
±1.25m/s
|
±2.50m/s
|
|
40~80°C
|
±0.35m/s
|
±0.70m/s
|
±1.75m/s
|
±3.50m/s
|
|
Selectable Output Option
|
(Set via onboard DIPswitch).
Current output: DC 4~20mA (Max. load resistance: 250Ω)
Voltage output: DC 0~5V
|
|
Selectable Range Option
|
(Rotary Switch selectable).
0~2, 0~5, 0~10, 0~25, or 0~50 m/s
|
|
Display Resolution (Model 6332D only)
|
Velocity Range: 0~2, 0~5, 0~10 m/s → Display Resolution: 0.01m/s
Velocity Range: 0~25, 0~50 m/s → Display Resolution: 0.1m/s
|
|
Power Supply
|
DC 12~24V
|
|
Power Consumption (Reference Value)
|
Approx. 2.0W (Under the following conditions - power supply: 12V, air velocity: approx. 10m/s, using a unit with a display with probe model 0965-03.)
Power consumption rate is subject to change according to conditions such as air velocity, probe type and use of display.
|
|
Connection Wire Size
|
0.5 ~ 1.5 mm2
|
|
Operating Temperature Range
|
5 ~ 40°C
|
|
Storage
Temperature Range
|
-10 ~ 50°C
|
|
Dimensions
|
Approx. 128(W)×78(D)×30(H) mm
|
|
Weight
|
Approx. 320g (11oz.)
|
|
Standard Accessory
|
Operation Manual : 1, Main Unit Case: 1
|
|
Optional Accessory
|
Probe Cable (10m, 20m, 30m), Display Unit (Supplied with a case with a display window and 2 installation screws.), Dedicated AC Adapter (DC12V)
|
Probe Specifications
|
MODEL
|
Measuring Range
|
Sensor Type
|
|
0962-00, 0963-00
|
0.1~50.0m/s
|
Uni-directional
|
|
0964-01, 0964-02
|
Omni-directional (Needle)
|
|
0965-00, 0965-01
|
0.1~25.0m/s
|
Omni-directional (Spherical)
|
|
0965-03, 0965-04
|
0.1~25.0m/s
|
Mini-temperature-compensation-sensor integrated type
Omni-directional (Spherical)
|
|
0965-07, 0965-08
|
Mini-temperature-compensation-sensor independent type
Omni-directional (Spherical)
|
|
|
|
|