Your cart is empty.



   
 

   

 

Martel - PTC-8010 RTD and Thermocouple Calibrator

<< Previous in Temperature Next in Temperature >>

Price: $1,599.99
Item Is In Stock
Item Number: PTC-8010 (#1920000)
Manufacturer: MARTEL
Mfr. Part Number: PTC-8001

RTD Calibrator and Thermocouple Calibrator

Precision Universal Temperature, RTD and Thermocouple Calibration Device
  • Simple, straight forward, Intuitive user interface
  • Easy to learn yet powerful 3 Key Martel Menu System
  • New ClearBrite™ Graphic Display
  • New streamlined case design
  • Calibrate thermocouples and RTDs with one unit
  • Works with Rosemount 3144 and all other smart transmitters
  • Direct keyboard entry or scroll control of output
  • User-defined setpoints
  • High accuracy ±0.4°C J thermocouple and ±0.3 °C PT 100-385 RTDs; all errors included
  • Thirteen (13) thermocouple types and thirteen (13) RTD types including PT 385 (1,000 Ohm) and Cu (10 Ohm)
  • NEMA 4 rated case, rugged design
  • RTD simulation works with all pulsed (smart) transmitters
  • YSI 400 series thermistor range for medical applications
  • Standard mini-plug connections for T/C
  • Standard banana jack connections for RTDs
  • Store setpoints for each output function
  • RS232 interface
  • Supplied in full rubber boot; optional carrying case available
  • Compare with other leading Manufacturers
  • Includes NIST Certificate with calibration data

Ranges: T/C: J,K,T,E,R,S,N,B,L,U,P including -10 to 75 mV; RTD: Pt 385 (100, 200, 500, 1000 Ohms), Pt 392, JIS, Ni120, Cu10, Cu50, Cu100, 4SI; Ohm: 0.0 to 400.0; 400 to 3,200; mV: -10 to +75
RTD IEX-Range: 0.01 mA to 3 mA
RTD Frequency Response: 10 ms; works with all pulsed transmitters
Environmental: Operating Temperature: -10C to +50C; Storage Temperature: -20C to 60C
Power Requirements: Batteries: 4 AA alkaline batteries, 30 hours type. op. time
Accessories Incl: Test leads, 4 AA batteries, NIST Certificate, instruction manual
 
Thermocouple Read and Source (all errors included)

Range Minimum Maximum Accuracy
J
-200.0
0.0
0.0
1,200.0
0.6C
0.4C
K
-200.00
0.0
0.0
1,370.0
0.8C
0.5C
T
-200.0
0.0
0.0
400.0
0.6C
0.5C 
E
-200.0
0.0
0.0
950.0
0.5C
0.4C 
R
-200.0
0.0
0.0
500.0
1,750.0 
2.4C
1.7C
1.3C
S
-200.0
0.0
0.0
500.0
1,750.0
2.4C
1.7C
1.3C 
B
-200.0
0.0
800.0
1,000.0
1,800.00 
2.1C
1.7C
1.3C 
L
-200.0
0.0
 0.0
900.0 
0.45C
0.4C
U
-200.0
0.0
 0.0
400.0
0.7C
0.45C 
N
-200.0
0.0
0.0
1,300.00 
1.1C
0.6C 
mV -10.0  75.0  0.014%+-2
 


RTD Read and Source

Range

Minimum

Maximum

Accuracy

NI120 (672)

-80.0

260.0

0.2C

Pt 100 (385)

-200.0

800.0

0.33C

Pt 100 (3926)

-200.0

630.0

0.3C

Pt 100 (3916)

-200.0

630.0

0.3C

Pt 200 (385)

-200.0

630.0

0.8C

Pt 500 (385)

-200.0

500.0

500.0

630.0

0.3C

0.4C

Pt 1000 (385)

-200.0

100.0

100.0

680.0

0.2C

0.3C

Cu10

-100.0

250.0

2.2C

YSI400

15.0

50.0

0.05C



View the Data/Spec Sheet. Don't have Reader?
View the Manual.