Features
    * Fast, portable and easy to use
    * Correct target size for most thermometers
    * Calibration solutions from –15 °C to 500 °C (5 °F to 932 °F)
    * Radio metrically calibrated for traceable and consistent results


Specifications


4180
4181
Temperature range
(@ 23 °C ambient,
0.95 emissivity)
–15 °C to 120 °C
(5 °F to 248 °F)
35 °C to 500 °C
(95 °F to 932 °F)
Display accuracy1
± 0.40 °C at –15 °C
± 0.40 °C at 0 °C
± 0.50 °C at 50 °C
± 0.50 °C at 100 °C
± 0.55 °C at 120 °C
± 0.35 °C at 35 °C
± 0.50 °C at 100 °C
± 0.70 °C at 200 °C
± 1.20 °C at 350 °C
± 1.60 °C at 500 °C
Stability
± 0.10 °C at –15 °C
± 0.05 °C at 0 °C
± 0.10 °C at 120 °C
± 0.05 °C at 35 °C
± 0.20 °C at 200 °C
± 0.40 °C at 500 °C
Uniformity2
(5.0 in dia of center
of target)
± 0.15 °C at –15 °C
± 0.10 °C at 0 °C
± 0.25 °C at 120 °C
± 0.10 °C at 35 °C
± 0.50 °C at 200 °C
± 1.00 °C at 500 °C
Uniformity2
(2.0 in dia of center
of target)
± 0.10 °C at -15 °C
± 0.10 °C at 0 °C
± 0.20 °C at 120 °C
± 0.10 °C at 35 °C
± 0.25 °C at 200 °C
± 0.50 °C at 500 °C
Heating time
15 min: –15 °C to 120 °C
14 min: 23 °C to 120 °C
20 min: 35 °C to 500 °C
Cooling time
15 min: 120 °C to 23 °C
20 min: 23 °C to –15 °C
100 min: 500 °C to 35 °C
40 min: 500 °C to 100 °C
Stabilization time
10 minutes
10 minutes
Nominal emissivity3
0.95
0.95
Thermometer emissivity compensation
0.9 to 1.0
Target diameter
152.4 mm (6 in)
Computer interface
RS-232
Power
115 V ac (± 10%), 6.3 A,
50/60 Hz, 630 W
230 V ac (± 10%), 3.15 A,
50/60 Hz, 630 W
115 V ac (± 10%), 10 A,
50/60 Hz, 1000 W
230 V a (± 10%), 5 A,
50/60 Hz, 1000 W
Fuse(s)
115 V ac 6.3 A, 250 V, slow blow
230 V ac 3.15 A, 250 V, T
115 V ac 10 A, 250 V, fast blow
230 V ac 5 A, 250 V, F
Size (HxWxD)
356 mm x 241 mm x 216 mm
(14 in x 9.5 in x 8.5 in)
356 mm x 241 mm x 216 mm
(14 in x 9.5 in x 8.5 in)
Weight
9.1 kg (20 lb)
9.5 kg (21 lb)
Safety
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04

1 - For 8 mm to 14 mm spectral band thermometers with emissivity set between 0.9 and 1.0
2 - The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the target will measure the same temperature.
3 - The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize emissivity related uncertainties.

   

Ordering Information

4180 - Precision Infrared Calibrator, –15 °C to 120 °C
4181 - Precision Infrared Calibrator, 35 °C to 500 °C
Included Accessories  - Accredited radiometric calibration report, target cover, User Guide, Getting Started Guide, and IR Configuration Utility
    



View the
4180/4181
Data/Spec Sheet
.
Don't have Reader?
View the 4180/4181
Manual
.
 

Related Products

4180-APRT - 2" Aperture, IR Instrument
Price: $530.00
4180-APRT - 2" Aperture, IR Instrument
2” Aperture, for 4180 or 4181 infra-red calibrator.
4180-CASE - Case, 4180-81, Carrying
Price: $735.00
4180-CASE - Case, 4180-81, Carrying
Case for models 4180-81 infra-red calibrators. Rugged, padded carrying or transport case.